R-1 to maintain 160 psig with 1,800 psig supply
pressure applied, proceed as follows:
CAUTION
VALVES V-2, V-5, V-6, V-7, AND V-10
ARE METERING (NEEDLE) VALVES.
OVERTIGHTENING THESE VALVES
WILL DAMAGE THE VALVE SEAT.
ONLY FINGERTIGHT PRESSURE
SHOULD BE USED WHEN YOU
CLOSE THESE VALVES.
1. Ensure that all test stand valves are closed,
and plug the bell jar bottom coupling C-1.
WARNING
WHEN YOU WORK WITH OXYGEN,
MAKE CERTAIN THAT CLOTHING,
TUBING FITTINGS, AND EQUIP-
MENT ARE FREE OF OIL, GREASE,
FUEL, HYDRAULIC FLUID, OR ANY
COMBUSTIBLE MATERIALS. FIRE
OR EXPLOSION MAY RESULT WHEN
EVEN SLIGHT TRACES OF COM-
BUSTIBLE MATERIAL COME IN
CONTACT WITH OXYGEN WHEN IT
IS UNDER PRESSURE.
2. Open the oxygen supply cylinder valve.
NOTE: When you set regulator R-1, a
MINIMUM of 1,800 psig of oxygen
pressure is applied to the regulator.
3. Open the test pressure gauge to bell jar
valve V-2 slowly, and fully open the oxygen supply
valve V-6.
4. Loosen the hex locknut located on the front
of regulator R-1. Turn the T-handle until 160 psig
registers on the test pressure gauge PG-1. Tighten
the hex locknut, and your oxygen pressure gauge
is now set.
5. Close the oxygen supply cylinder valve and
open the system bleed valve V-5 to bleed pressure
from the system. After bleeding the pressure,
remove the plug from the bell jar bottom coupling
C-1.
LEAKAGE TEST, ACCESSORIES
SECTION. TO perform the leakage test on the
accessories section of the test stand, proceed as
follows:
1. Install the nipple assembly in the bell
jar bottom coupling C-1. Connect one end of the
hose to the adapter, and the other end to the
differential pressure connection NIP-7.
2. Ensure that the test pressure gauge to bell
jar valve V-2 is open. The system bleed valve V-5,
the test pressure gauge buildup, and the vent valve
V-10 and the differential pressure bleed valve V-7
are closed.
3. Open the differential pressure shutoff valve
V-8 and the oxygen supply cylinder valve.
4. Slowly open the oxygen supply valve V-6
until 160 psig is indicated on test pressure gauge
PG-1.
5. Now close the oxygen supply valve V-6.
Leakage will be indicated by a drop in pressure
on PG-1. Leakage should not be more than 2 psig
in 10 minutes.
6. Leave all hoses and valves in their present
position and start your test stand leakage test.
LEAKAGE TEST, TEST STAND. T O
perform the leakage test on the entire test stand
proceed as follows:
1. Open the converter supply flow control
valve V-9 and test pressure gauge buildup and the
flow valve V-10.
2. Plug the converter supply outlet NIP-5 and
the supply converter connection NIP-6. Ensure
that the system bleed valve (V-5) is closed.
3. Open the supply valve V-6 until the relief
valve RV-11 unseats. (The relief valve is set to
relieve at approximately 110 psig, and be leaktight
at 100 psig.) Using the system bleed valve V-5,
decrease pressure until 100 psig is indicated on test
pressure gauge PG-1. Close valve V-6. Leakage
will be indicated by a drop in pressure on PG-1.
Leakage should be no more than 10 psig in 10
minutes.
4. Bleed the test standby opening the system
bleed valve (V-5). Close all the test stand valves.
Remove the plugs from the converter supply outlet
NIP-5 and the plug from the supply converter
connection (NIP-6).
BELL JAR ASSEMBLY LEAKAGE
TEST. To perform a leakage test on the bell jar
assembly, proceed as follows:
1. Remove the hose assembly and the nipple
assembly from the bottom bell jar coupling C-1.
Disconnect the opposite end of the hose from
differential pressure connection NIP-7.
11-19
